E762: Alarm "NO X-ray feedback"

Updated: 06/16/2017
Article #: 93


This message is displayed when the Generator CPU board (A9) is not
detecting the X-ray emission feedback signal.

 

In this case, check the correct insertions of X53-X53 (flat cable) between
Generator board and Generator CPU board (A9).

 

Check the CANBus connection between the Generator CPU board (A9)
and the main CPU A5 (cable X20-X20).

 

Check the correct insertion of X56 and X57 between the Generator board
to the tubehead.

 

If all connections are OK and the fault is repetitive, replace the Generator
CPU board (A9).

 

 