E756: PFC failure

Updated: 06/16/2017
Article #: 90


This message is signalling that the PFC (Power Factor Control) circuit is
not correctly functioning.

 

If the error is present in standby mode, replace the Generator board
(A10).
If the error appears during an exposure, reset the error and repeat the
exposure; if the error still appears, replace the Generator board and then
the tubehead.

 

 