Errors with code from E400 to E499
|
|||||
Updated: 06/16/2017
Article #: 84
|
|||||
E400: Timeout of zero position sensor on Soft Tissue Filter /
The Soft Tissue Filter for cephalometric examinations is controlled by the
• Power off the system and remove the frontal cover of the tubehead
• Check that during the power on phase there are movements on the
• Check continuity of the above cables, from M6 to X65 and from X65
• Repeat the power on sequence, verifying the motion.
• Power off the system and try to manually disengage the sensor,
• Power on the system and check the functioning of the sensor,
• If the variation is observed and the message still continues to be
• If all cables are OK, the fault is on the CPU, so replace it.
|
|||||
|
|||||
|
|||||
|