E340: Sensor holder not in PAN position

Updated: 06/16/2017
Article #: 79


An examination requiring the sensor holder in PAN position has been
selected, but it is not sensed in the proper position.

 

Close the PAN sensor holder to its position and press the Patient
centering key to restart the power on procedure.
If this does not reset the error message:

 

• Check the proper function of optical sensor B6.
• Check continuity of cables from optical sensor to A8 board, connector
X48, and from that board to main CPU board (A5), connector X33.