How To: EVO TS Twain Module Spins
|
|||||
Updated: 02/27/2018
Article #: 131
|
|||||
See video HERE for how to fix the issue
Step 1: rename the Villa Sistemi Medicali in C:\ProgramData\Villa Sistemi Medicali Step 2: check CPU / TSE firmware from master control of Pano (see Villa logo, software revisions) Step 3: firmware of the machine will determine appropriate twain driver (here CPU FW 505 / TSE 103 will require Twain 505) Step 4: Source the twain package, and run the installation Step 5: make sure to copy the calibration data into the new folder created after the install Step 6: open the module making sure of the connection and the fix Step 7: double check the appropriate filtering plugin is applied, and if not, cancel of out of the capture window, select the drop down and hit save
Note: This issue is not related to USB dongle key, occurring with and without key enabled |
|||||
|
|||||
|
|||||
|